Features:
*Advanced optical path design: SKZ1062C-2 adopts concentrated light Fourier transform test technology to ensure the maximum range in the shortest focal length, effectively improve the resolution of the instrument; unique high-density detection unit, so that SKZ1062C-2 has super small particles Test capability, high-density detection unit uses the SKZ1062C-2 with superior full-scale seamless testing capability.
* Single-lens single-lens design: The newly designed optical path adopts the latter lens acquisition technology, which realizes no blind zone detection from 0.01 to 170 degrees, and the measurement range is from 0.01 micron to 2000 micron.
*Aluminum alloy optical platform: high-precision all-aluminum optical platform, long-term stable and reliable; optical path system is fully enclosed, effectively solve the dust pollution problem; the instrument design meets the international laser safety regulations, with laser beam automatic protection.
*Operational protection function: The newly designed SKZ1062C-2 has misoperation protection function. The instrument has non-response function to the operator's wrong command, which avoids the damage of the instrument caused by misoperation and greatly improves the stability of the instrument test result.
* Large-caliber lens: Large-caliber Fourier transform lens, more efficient implementation of optical signal Fourier transform, more accurate and more effective.
* Ultrasonic anti-dry: SKZ1062C-2 realizes the waterless ultrasonic self-protection function of the sample cell to avoid damage to the ultrasound; at the same time, the ultrasonic power can be adjusted within the range of 0 to 50 watts.
Dustproof and shockproof design: The whole instrument has been sealed and designed to greatly improve the service life of internal components. The unique suspension structure can effectively avoid the interference of external vibration to the
instrument, and make the result test more stable and reliable.
* Strong anti-corrosion design (optional): It can be equipped with acid, alkali, oil (including all solvent oil) and organic
solvent (such as acetone, phenol, n-hexane and other organic solvents) according to the actual needs of customers.
*Imported fiber-optic semiconductor laser: SKZ1062C-2 adopts high-stability, long-life imported high-power fiber-optic output semiconductor laser. Excellent monochromaticity and stability make SKZ1062C-2 have excellent test repeatability; laser has self-checking The power change self-regulation function keeps the laser constant power output. (wavelength 635 nm; power greater than 5 mw).
*Automatic alignment: The exclusive use of the combination of mechanical center and optical center, the optical path automatically adjusts the positioning more accurately, reaching the micron level; at the same time, the optical path adjustment speed is faster and faster and can be adjusted in 15 seconds. The self-developed automatic alignment system consists of stepper motor, precision guide rail, precision controller and software system. The minimum step distance is 0.2 microns, ensuring that the focus of the laser beam always passes through the center point of the detector, improving the accuracy of the test results and testing of Repeatability.
* Split design: SKZ1062C-2 adopts split design, the optical path part is more stable and effective, avoids the influence of the vibration of the dispersing device, and greatly improves the test repeatability of the instrument.
* Free bubble design: The new design makes no bubbles enter the test sample window during the whole test process, avoiding air bubble interference.
* Sample no residue design: The instrument pipeline and drainage structure have been optimized. There is no residual liquid in the instrument pipeline and circulating pump to avoid the impact on the next test data. The dry test also has no residue design.
* Dry computer remote control feeding: When SKZ1062C-2 dry test, testers can control the feeding speed through the computer remotely, which greatly reduces the labor intensity of testers.
* Sample window quick change device: The newly designed sample window quick change device makes the sample window replacement more convenient and quick.
*Software: Compliance with Pharmacopoeia GMP regulations, with electronic signature, permission settings, data tracking, data can not be changed and other functions.
Specification
Standard |
GB/T 29022-2012, ISO 22412 |
Test Range |
1--10000nm |
Concentration range |
0.1mg/L --100mg/L |
Accuracy |
<1% |
Repeatability error |
<1% |
Laser |
Main laser: λ=532nm, semiconductor laser (temperature control protection), P>30mW |
Detector |
Photomultiplier tube (PMT) |
Scattering angle |
90° |
Sample cell |
10mm*10mm , 4ml |
Temperature range |
5-90 ° C (accuracy ± 0.1 ° C) |
Rated voltage |
AC110V/ AC220V |
Test speed |
<1Min/time (excluding sample dispersion time) |
Photon correlator |
Physical channel: 512, equivalent channel: 10000, baseline channel: 8 Sampling and delay time: 1us~200us dynamically adjustable Minimum resolution: 6ns |