Customization: | Available |
---|---|
Warranty: | 12 Months |
Usage: | Textile Fabric Testing |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Used for testing the filtered out percentage of bacterium-containing suspended particles (BFE) of medical and daily protective mask materials under specified flow rate.
Standards
YY0469 ASTMF2100 ASTMF2101 EN14683
Feature
Main parameters | Specification | Resolution | Max. allowable error |
Sample flow path | 28.3L/min | 0.1L/min | ±2% |
Spray flow | (8-10)L/min | 0.1L/min | ±5.0% |
Peristaltic pump flow | (0.006~3.0)mL/min | 0.001ml/min | ±2.% |
A channel pressure flow meter | (-20-0)kPa | 0.01kPa | ±2.5% |
Sample flowmeter front pressure | (-20-0)kPa | 0.01kPa | ±2% |
Spray flowmeter pressure | (0-300)kPa | 0.1kPa | ±2% |
Aerosol chamber negative pressure | (-90- -120)Pa | 0.1Pa | ±1.0% |
Operating temperature | 0-50ºC | ||
Cabinet negative pressure | ≥120Pa | ||
Data storage capacity | Expandable | ||
High efficiency air filter performance | ≥99.995%@0.3μm,≥99.9995%@0.12μm | ||
Aerosol generator mass median diameter | Average particle diameter (3.0±0.3)μm Geometric standard deviation ≤1.5 |
||
Dual 6-stage Anderson sampler captures particle size | Grade1> 7μm, Grade 2 (4.7-7)μm, Grade3 (3.3-4.7)μm, Grade 4 (2.1-3.3)μm, Grade 5 (1.1-2.1)μm, Grade 6 (0.6-1.1)μm | ||
Total number of positive QC sampler particles | (2200±500)cfu | ||
Negative pressure cabinet ventilation flow | ≥5m3/min | ||
Negative pressure cabinet size | 600x85x3mm | ||
Host size | 1000x600x690mm | ||
Bracket size | 1470x790x2100mm | ||
Power | AC220V±10%,50Hz 1kW | ||
Noise | <65db |